Objectives: To determine the assigned values and evaluate the Z-score performance of laboratories participating in an external quality assessment pilot study for coagulation testing.
Methods: Coagulation external quality assessment samples at 3 concentration levels were produced and distributed to 67 participating laboratories in Ho Chi Minh city, Hue, and the Mekong Delta region. Robust statistical methods, in accordance with ISO 13528:2022, were employed to establish assigned values for PT, APTT, and fibrinogen based on participant data. The study further assessed the percentage of laboratories achieving acceptable Z-score across three major instrument platforms: Sysmex CS series, Stago Compact Max, and ACL TOP series.
Results: The assigned values for PT and APTT exhibited variations exceeding 10%, primarily due to methodological discrepancies between different instrumentation platforms. Regarding Z-score performance, 100% of laboratories achieved acceptable scores for fibrinogen, while the success rates for APTT and PT were over 85% and 70%, respectively.
Conclusions: This study provides a comprehensive overview of the current status of laboratory performance, serving as a basis for quality monitoring, training, and continuous improvement. The findings establish a significant foundation for the broader implementation of coagulation external quality assessment programs in the near future.
Article Details
Keywords
External quality assessment, quality assurance, coagulation testing, assigned value, Z-score.
References
[2] Clinical and Laboratory Standards Institute (CLSI). Collection, transport, and processing of blood specimens for testing plasma-based coagulation assays and molecular hemostasis assays, 7th ed, 2021.
[3] World Health Organization. Laboratory quality management system, 2nd ed, Geneva, 2021.
[4] Bakaltcheva I, O’Sullivan A.M, Hmel P et al. Freeze-dried whole plasma: evaluating sucrose, trehalose, sorbitol, mannitol and glycine as stabilizers. Thromb Res, 2007, 120 (1): 105-116.
[5] International Organization for Standardization. ISO 13528: Statistical methods for use in proficiency testing by interlaboratory comparison, 3rd ed, 2022.
[6] United States Government Publishing Office. Clinical Laboratory Improvement Amendments (CLIA) Regulations, Part 493: Laboratory Requirements. Washington, DC, 2024. p. 7562-7615.
[7] Trần Hữu Tâm. Ngoại kiểm tra chất lượng xét nghiệm (tái bản lần 4). Nhà xuất bản Y học, 2020.
[8] Trần Minh Điển, Cao Việt Tùng, Phan Hữu Phúc và cộng sự. Đánh giá độ tương đồng kết quả xét nghiệm một số chỉ số đông máu cơ bản trên hệ thống xét nghiệm đông máu tự động ACL TOP 750 LAS và STAGO STA R Max. Tạp chí Y Dược lâm sàng 108, 2025, 20 (2):139-146. doi: 10.52389/ydls.v20i2
[9] Geens T, Vertessen F, Malfait R et al. Validation of the Sysmex CS 5100 coagulation analyzer and comparison to the Stago STA‐R analyzer for routine coagulation parameters. International Journal of Laboratory Hematology, 2015, 37 (3): 372-381. doi: 10.1111/ijlh.12295
[10] Trần Hữu Tâm. Đánh giá chất lượng của 18 thông số xét nghiệm đông máu tại các phòng xét nghiệm. Tạp chí Y học Việt Nam, 2022, 514 (2): 107-110. doi: 10.51298/vmj.v514i2.2608