EFFECTS OF HEMOLYSIS ON SELECTED BLOOD BIOCHEMISTRY TEST RESULTS AND IMPLICATIONS FOR LABORATORY QUALITY AT TAM DAO REGIONAL MEDICAL CENTER, PHU THO PROVINCE, VIETNAM, IN 2026

Nguyen Van Vinh, Tran Anh Huy, Bui Thanh Son, Pham Quynh Trang, Pham Thu Phuong, Tran My Linh, Vu Dinh Tu
DOI: 10.52163/yhc.v67iE2.7202
0 Views
0 Downloads
Abstract

Objectives: To compare selected biochemical parameters in paired hemolyzed and non-hemolyzed specimens and discuss implications for laboratory quality at Tam Dao Regional Medical Center in 2026.

Methods: A descriptive cross-sectional study with paired-specimen comparisons included all eligible hemolyzed specimens collected between April 1 and October 1, 2026. Each hemolyzed specimen was paired with a non-hemolyzed specimen collected from the same patient within 30 minutes after hemolysis was detected. Glucose, creatinine, aspartate aminotransferase (AST), alanine aminotransferase (ALT), potassium, and total bilirubin were assessed. Data were analyzed using Microsoft Excel.

Results: The study included 21 specimen pairs. Hemolysis grades I, II, and III accounted for 52.38%, 33.33%, and 14.29% of pairs, respectively. Glucose and creatinine tended to be slightly lower, ALT higher, and total bilirubin lower in hemolyzed specimens. AST and potassium showed the largest increases, with differences tending to increase across hemolysis grades. Reported p-values indicated statistically significant differences for AST in all three grades, potassium in grades I and II, creatinine in grade I, and total bilirubin in grade III. The potassium difference in grade III was not statistically significant.

Conclusion: In this small, single-center paired-specimen study, hemolyzed specimens showed higher AST and potassium measurements than non-hemolyzed specimens. These findings support attention to specimen quality and analyte-specific reporting decisions but do not establish validated rejection thresholds or the effectiveness of quality-improvement interventions.

References
[1]
[1]. Nordin N, Ab Rahim SN, Wan Omar WFA, Zulkarnain S, Sinha S, Kumar S, et al. Preanalytical Errors in Clinical Laboratory Testing at a Glance: Source and Control Measures. Cureus. 2024;16(3):e57243. doi:10.7759/cureus.57243. Google Scholar
[2]
[2]. Yeganeh BS, Soleimani N, Zare S, Mohammadzadeh S, Amiri F, Esmaeili MJ, et al. The Effect of Hemolysis on Biochemical Tests: Determining Interference Cutoffs and Managing Hemolytic Samples. Journal of Analytical Methods in Chemistry. 2026;2026:9307894. doi:10.1155/jamc/9307894. Google Scholar
[3]
[3]. Glick MR, Ryder KW, Glick SJ, Woods JR. Unreliable visual estimation of the incidence and amount of turbidity, hemolysis, and icterus in serum from hospitalized patients. Clinical Chemistry. 1989;35(5):837–839. doi:10.1093/clinchem/35.5.837. Google Scholar
[4]
[4]. Ryder KW, Glick MR, Glick SJ. Incidence and Amount of Turbidity, Hemolysis, and Icterus in Serum From Outpatients. Laboratory Medicine. 1991;22(6):415–418. doi:10.1093/labmed/22.6.415. Google Scholar
[5]
[5]. Wan Azman WN, Omar J, Koon TS, Tuan Ismail TS. Hemolyzed Specimens: Major Challenge for Identifying and Rejecting Specimens in Clinical Laboratories. Oman Medical Journal. 2019;34(2):94–98. doi:10.5001/omj.2019.19. Google Scholar
[6]
[6]. Trung tâm Y tế khu vực Tam Đảo. Quyết định số 97/QĐ-TTYTKV về việc phê duyệt đề cương và triển khai thực hiện đề tài nghiên cứu khoa học năm 2026. 2026. Google Scholar
[7]
[7]. Koseoglu M, Hur A, Atay A, Cuhadar S. Effects of hemolysis interferences on routine biochemistry parameters. Biochemia Medica. 2011;21(1):79–85. doi:10.11613/bm.2011.015. Google Scholar