Articles Tập 67 Số English Version No 2, 2026 27/11/2026

ESTABLISHMENT OF REFERENCE INTERVALS FOR 10 ROUTINE BIOCHEMICAL ANALYTES IN APPARENTLY HEALTHY VIETNAMESE ADULTS USING DIRECT SAMPLING

Nguyen Hoang Bac, Mai Thi Bich Chi, Tong Nguyen Thanh Cong, Vo Phi Gia Bao, Nguyen Chi Vinh, Nguyen Thi Bang Suong
DOI: 10.52163/yhc.v67iE2.7200
0 Views
0 Downloads
Abstract

Objective: To establish reference intervals (RIs) for 10 routine biochemical analytes in healthy Vietnamese adults using the Beckman Coulter AU5800 analyzer.

Methods: This cross-sectional study used direct a priori sampling of healthy adults aged 18–60 years following CLSI EP28-A3c. Specimens were analyzed on a Beckman Coulter AU5800 analyzer. Partitioning was assessed using the Harris–Boyd test, and RIs were estimated using nonparametric or robust methods with 90% confidence intervals. Sensitivity and age-adjusted analyses were performed.

Results: The reference population comprised 280 adults (85 men and 195 women). Eight analytes met the criteria for sex-specific RIs (men; women): glucose, 4.40–5.66; 3.80–5.53 mmol/L; urea, 3.22–7.53; 2.70–6.21 mmol/L; creatinine, 69.5–109.5; 54.1–83.0 µmol/L; high-density lipoprotein cholesterol (HDL-C), 0.95–1.90; 1.00–2.11 mmol/L; triglycerides, 0.48–3.41; 0.43–2.07 mmol/L; aspartate aminotransferase (AST), 16.6–40.5; 14.0–28.0 U/L; alanine aminotransferase (ALT), 8.2–47.8; 5.5–24.9 U/L; and gamma-glutamyl transferase (GGT), 14.7–96.7; 9.0–34.8 U/L. Total cholesterol had a combined-sex RI of 3.60–6.12 mmol/L. Low-density lipoprotein cholesterol (LDL-C) had combined-sex, age-specific RIs of 2.00–3.83 mmol/L for ages 18–35 years and 2.18–4.22 mmol/L for ages 36–60 years. Sex differences persisted after age adjustment; the female ALT upper limit was sensitive to outlier exclusion.

Conclusion: Eight analytes required sex-specific RIs, and LDL-C required age-specific RIs. The female ALT upper limit was lower than the combined-sex limits reported in two Vietnamese studies but depended on outlier exclusion and should be considered provisional. These RIs require local verification; the small male sample and limited representation of older adults constrain their precision.

References
[1]
1. Kilickaya O, Schmickl C, Ahmed A, Pulido J, Onigkeit J, Kashani K, et al. Customized reference ranges for laboratory values decrease false positive alerts in intensive care unit patients. PLoS One. 2014;9(9):e107930. doi: 10.1371/journal.pone.0107930 Google Scholar
[2]
2. Clinical and Laboratory Standards Institute. Defining, establishing, and verifying reference intervals in the clinical laboratory; approved guideline – third edition. CLSI document EP28-A3c. Wayne (PA): Clinical and Laboratory Standards Institute; 2010. Google Scholar
[3]
3. Harris EK, Boyd JC. On dividing reference data into subgroups to produce separate reference ranges. Clin Chem. 1990;36(2):265-70. doi: 10.1093/clinchem/36.2.265 Google Scholar
[4]
4. Le AN, Nguyen KT, Le DV, Nguyen TV, Tran TV, Nguyen TN, et al. [Study on establishing reference intervals for selected biochemical and hematological parameters at Binh Thanh District Health Center in 2024]. Vietnam Journal of Science and Technology – B. 2026. Online first. Vietnamese. doi: 10.31276/VJST.2026.3909 Google Scholar
[5]
5. Hoang VT, Than QP, Vu TH, Nguyen TT, Quach TH, Tran KH, et al. [Establishing reference intervals for selected cellular hematological and biochemical parameters in adults]. Hong Bang International University Journal of Science. 2026;(42):31-40. Vietnamese. doi: 10.59294/HIUJS2026086 Google Scholar
[6]
6. Wang D, Ma C, Zou Y, Yu S, Li H, Cheng X, et al. Gender and age-specific reference intervals of common biochemical analytes in Chinese population: derivation using real laboratory data. J Med Biochem. 2020;39(3):384-91. doi: 10.2478/jomb-2019-0046 Google Scholar
[7]
7. Xia L, Chen M, Liu M, Tao Z, Li S, Wang L, et al. Nationwide multicenter reference interval study for 28 common biochemical analytes in China. Medicine (Baltimore). 2016;95(9):e2915. doi: 10.1097/MD.0000000000002915 Google Scholar
[8]
8. Ichihara K, Ceriotti F, Tam TH, Sueyoshi S, Poon PMK, Thong ML, et al. The Asian project for collaborative derivation of reference intervals: (1) strategy and major results of standardized analytes. Clin Chem Lab Med. 2013;51(7):1429-42. doi: 10.1515/cclm-2012-0421 Google Scholar
[9]
9. Kwo PY, Cohen SM, Lim JK. ACG clinical guideline: evaluation of abnormal liver chemistries. Am J Gastroenterol. 2017;112(1):18-35. doi: 10.1038/ajg.2016.517 Google Scholar