Objective: This study aims to assess the safety and effectiveness of using anti-inflammatory and
bactericidal plasma generated from the PlasmaMed machine manufactured in Vietnam for treating
infected wounds.
Methods: This study was a controlled clinical trial involving 31 patients with infected wounds.
Of the 31 patients, 20 had two similar wound areas, with one area treated with plasma irradiation
while the other received routine dressing changes and served as the control group. The remaining 11
patients had only one wound area treated with plasma irradiation. The efficacy of the treatment was
measured using various indicators, including C-reactive protein (CRP), procalcitonin (PCT), and
bacterial concentration at the wound site.
Results: The study showed a significant decrease in CRP and PCT levels immediately after the first
and fifth day of cold plasma irradiation. Additionally, the concentration of bacteria on the wound
surface was significantly reduced after the first plasma irradiation. By day 5, most of the patients’
wounds were completely clean, with positive changes observed, such as fast atrophy, reduced
exudation, healing of wound edges, and tissue growth, resulting in a uniform coloration of the overall
wound surface.
Conclusion: Based on the study results, it can be concluded that cold plasma generated from the
PlasmaMed machine has anti-inflammatory and bactericidal effects that can prevent infection in soft
wounds immediately after the first exposure and up to five days after treatment.
9. STUDY ON EFFECTIVENESS OF ANTI-INFLAMMATORY, ANTI-MICROBIAL OF COLD PLASMA IN PATIENTS WITH INFECTIOUS WOUND
18
Views
7
Downloads
Keywords
PlasmaMed, cold plasma, technology, wound care.
Abstract
References
[1]
Nguyễn Đình Bảng, “Phòng chống nhiễm khuẩn
Google Scholar
[2]
tại các cơ sở ngoại khoa”. Tạp chí Y học Việt
Google Scholar
[3]
Nam, (5), tr. 27 – 31, 1991.
Google Scholar
[4]
Tian Y et al., “Inactivation of Staphylococcus
Google Scholar
[5]
aureus and Enterococcus faecalis by a directcurrent,
Google Scholar
[6]
cold atmospheric-pressure air plasma
Google Scholar
[7]
microjet”, J Biomed Res. 24(4), 264-9, 2010.
Google Scholar
[8]
Kong MG et al., “Plasma medicine: an
Google Scholar
[9]
introductory review”, New Journal of Physics.
Google Scholar
[10]
(11), 115012, 2009.
Google Scholar
[11]
Boekema BKHL, Vlig M, Guijt D et al., A new
Google Scholar
[12]
flexible DBD device for treating infected wounds:
Google Scholar
[13]
in vitro and ex vivo evaluation and comparison
Google Scholar
[14]
with a RF argon plasma jet. J Phys D Appl Phys.
Google Scholar
[15]
;49:044001, 2016.
Google Scholar
[16]
Boekema BKHL, Hofmann SS, van Ham BJT et
Google Scholar
[17]
al., Antibacterial plasma at safe levels for skin
Google Scholar
[18]
cells. J Phys D Appl Phys. 46:422001, 2013.
Google Scholar
[19]
Morfill G, Shimizu T, Steffes B et al.,
Google Scholar
[20]
Nosocomical infections-a new approach towards
Google Scholar
[21]
preventive medicine using plasmas. New J Phys.
Google Scholar
[22]
;11:115019, 2009.
Google Scholar
[23]
Li YF, Taylor D, Zimmermann JL, In vivo skin
Google Scholar
[24]
treatment using two portable plasma devices:
Google Scholar
[25]
comparison of a direct and an indirect cold
Google Scholar
[26]
atmospheric plasma treatment. Clin Plasma Med.
Google Scholar
[27]
;1:35–9, 2013.
Google Scholar
[28]
Do Hoang Tung, Cold atmospheric pressure
Google Scholar
[29]
gliding arc plasma jet for decontamination,
Google Scholar
[30]
Communications in Physics, Vol. 24, No. 3S2,
Google Scholar
[31]
pp. 129-134, 2014.
Google Scholar
[32]
Isbary G, Morfill G, Schmidt HU et al., A first
Google Scholar
[33]
prospective randomized controlled trial to
Google Scholar
[34]
decrease bacterial load using cold atmospheric
Google Scholar
[35]
argon plasma on chronic wounds in patients, Br J
Google Scholar
[36]
Dermatol 163(1), 78-82, 2010.
Google Scholar