THE AGREEMENT OF SELECTED PERIPHERAL BLOOD CELL PARAMETERS BETWEEN THE AUTOMATED HEMATOLOGY ANALYZERS DXH 900 AND XN 10 AT NATIONAL INSTITUTE OF HEMATOLOGY AND BLOOD TRANSFUSION

Truong Thi Hien1,2, Nguyen Ngoc Dung1,2
1 National Institute of Hematology and Blood Transfusion
2 Viện Huyết học- Truyền máu Trung ương
DOI: 10.52163/yhc.v67iCD11.6043
20 Views
0 Downloads
Abstract

Objectives: Comparison of the results of Several Peripheral Blood Cell Analysis Tests on the DXH 900 and XN-10 automated hematology system at the National Institute of Hematology and Blood Transfusion in 2025.

Methods: A cross-sectional method. 80 blood specimens. Samples were analyzed on both DxH 900 and XN 10 analyzers to compare white blood cell count (WBC), platelet count (PLT), red blood cell count (RBC), hemoglobin (HGB), and mean corpuscular volume (MCV), following the CLSI EP09-C guidelines.

Results: All evaluated parameters (WBC, PLT, RBC, HGB, and MCV) demonstrated strong correlation (R > 0.99). Outcome A was observed for PLT (bias −2.4%) and RBC (bias 0.1%) between the two analyzers. Outcome B was identified for WBC (bias −2.0%), HGB (bias 3.4%), and MCV (bias 0.6%). Intraclass correlation coefficients (ICC) for all parameters exceeded 0.9.

Conclusion: The WBC, PLT, RBC, HGB, and MCV tests performed on the DXH 900 and XN 10 analyzers demonstrated good agreement.

References
[1]
R. D. Nerenz, M. E. Pittman, and M. Scott, "Impact of errors and variability on clinical laboratory test interpretation," in Pathobiology of Human Disease: A Dynamic Encyclopedia of Disease Mechanisms, ed Elsevier Inc., 2014, pp. 3222-3236. DOI:10.1016/B978-0-12-386456-7.06303-6 Google Scholar
[2]
J. Budd, A. Durham, T. Gwise, D. Hawkins, M. Holland, B. Iriarte, et al., "CLSI EP09-C, Measurement Procedure Comparison and Bias Estimation using Patient Samples– Approved Guideline," 2018. Google Scholar
[3]
ISO 15189:2022, "Medical laboratories — Requirements for quality and competence, International standard," ed, 2022. Google Scholar
[4]
E. E. F. o. C. C. a. L. Medicine. Available: https://biologicalvariation.eu/ Google Scholar
[5]
T. K. Koo and M. Y. Li, "A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research," J Chiropr Med, vol. 15, pp. 155-63, Jun 2016. DOI: 10.1016/j.jcm.2016.02.012 Google Scholar
[6]
J. Wang, S. Zhao, Z. Su, and X. Liu, "Analytical comparison between two hematological analyzer systems: Mindray BC-5180 vs Sysmex XN-1000," J Clin Lab Anal, vol. 33, p. e22955, Oct 2019. DOI: org/10.1002/jcla.22955 Google Scholar
[7]
M. Serrando Querol, J. Nieto-Moragas, A. Marull Arnall, M. D. Figueras, and O. Jiménez-Romero, "Evaluation of the new Beckmann Coulter analyzer DXH 900 compared to SYSMEX XN20: analytical performance and flagging efficiency," Diagnostics, vol. 11, p. 1756, 2021. DOI: 10.3390/diagnostics11101756 Google Scholar
[8]
K. Fujimaki, K. Hummel, I. Magonde, K. Dammert, Y. Hamaguchi, K. Mintzas, et al., "Performance evaluation of the new Sysmex XR-Series haematology analyser," Practical laboratory medicine, vol. 39, p. e00370, 2024. DOI: 10.1016/j.plabm.2024.e00370 Google Scholar