Articles Vol. 66 No. CĐ1-NCKH 24/01/2025

7. COMPARISON OF RAPID CENTRIFUGATION WITH ROUTINE CENTRIFUGATION FOR SOME COAGULATION TESTING (PT, APTT, FIBRINOGEN) AT DEPARTMENT OF HEMATOLOGY IN VIETDUC UNIVERSITY HOSPITAL

Nguyen Thu Hanh1,2, Ha Van Phu1,2, Tran Thi Hang1,2, Doan An Son1,2, Nguyen Van Chinh1,2, Nguyen Manh Cuong1,2, Nguyen My Van1,2
1 Vietduc University Hospital
2 Bệnh viện Hữu nghị Việt Đức
Corresponding author: hanhhhtm@yahoo.com
DOI: 10.52163/yhc.v66iCD1.1968
87 Views
85 Downloads
Abstract

Objectives: Comparison of rapid centrifugation (2300g in 5 minutes) with routine centrifugation (1500g  in 15 minutes) for some coagulation testing (PT, APTT and fibrinogen) at Department of Hematology in Vietduc University Hospital.

Methods: Experimental study on 45 pairs of patient samples according to the Clinical and Laboratory Standards Institute (CLSI) guideline for Method Comparison and Bias Estimation Using Patient Samples (EP09c).

Results: Platelet count after rapid centrifugation was 3.53 ± 1.50 G/l, similar to routine centrifugation (3.45 ± 1.58 G/l) with no statistically significant difference (p > 0.05). The results of Passing-Bablok linear regression analysis and Bland-Altman difference plot estimation showed that the difference in PT-INR and APTT-r and fibrinigen between the two centrifugation methods was within acceptable limits.

Conclusions: The present study showed no no statistically significant difference when comparable results between rapid (2300g × 5 minutes) and routine (1500g × 15 minutes) centrifugation techniques for some coagulation screening test (PT and APTT and fibrinogen). Rapid centrifugation can be used in emergency coagulation screening tests, especially in red code situations.

References
[1]
Australasian College for Emergency Medicine, Guidelines on the Implementation of the Australasian Triage Scale in Emergency Departments, 2023. Google Scholar
[2]
Bộ Y tế, Hướng dẫn chẩn đoán và điều trị bệnh lý huyết học, 2022. Google Scholar
[3]
CLSI, H21-A5: Collection, Transport, and Processing of Blood Specimens for Testing Plasma-Based Coagulation Assays and Molecular Hemostasis Assays, Approved Guideline - Fifth Edition, 2008. Google Scholar
[4]
Bệnh viện Hữu nghị Việt Đức, Quy trình báo động đỏ nội viện (Code Red), 2018. Google Scholar
[5]
CLSI, EP09cED3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples, Thirt edition, 2008. Google Scholar
[7]
Sultan A, Five-minute preparation of platelet-poor plasma for routine coagulation testing, East Mediterr Health J, 2010, 16 (02), 233-236. Google Scholar
[8]
Carroll W.E, Wollitzer A.O, Harris L et al, The significance of platelet counts in coagulation studies, J. Med, 2001, 32 (1-2), 83-96. Google Scholar
[9]
Chandler E, Kakkar, Comparison of Rapid Centrifugation Technique with Conventional Centrifugation for Prothrombin Time (PT) and Activated Partial Thromboplastin Time (APTT) Testing, Indian J. Hematol Blood Transfus, 2019, 35 (1), 161-166. Google Scholar
[10]
Quirke W, Koohestani Z, Revisiting coagulation centrifugation protocol for integration into total laboratory automation workflow, Journal of Laboratory and Precision Medicine, 2020, 5 (0). Google Scholar
[11]
Boissier E, Sévin-Allouet M, A 2-min at 4500g rather than a 15-min at 2200g centrifugation does not impact the reliability of 10 critical coagulation assays, Clinical Chemistry and Laboratory Medicine (CCLM), 2017, 55 (6), e118-e121. Google Scholar
[12]
Suchsland J, Friedrich N, Grotevendt A et al, Optimizing centrifugation of coagulation samples in laboratory automation, Clinical Chemistry and Laboratory Medicine (CCLM), 2014, 52 (8), 1187-1191. Google Scholar